INTOUCH® CRITICAL CARE BED - (CONTINUED)
Recommended separation distances between portable and mobile RF
communications equipment and the InTouch® Critical Care Bed.
The InTouch® Critical Care Bed is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled . The customer or the user of the InTouch® Critical Care Bed can help
prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the InTouch® Critical Care Bed as recommended below,
according to the maximum output power of the communications equipment .
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer .
Note 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies .
Note 2
These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people .
www.stryker.com
EMC Information
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d=1,2√P
0,12
0,38
1,2
3,8
12
2151-109-006 REV A
m
800 MHz to 2,5 GHz
d=1,2√P
0,12
0,38
1,2
3,8
12
Return To Table of Contents
d=2,3√P
0,23
0,73
2,3
7,3
23
1-85