CONFIGURATION SCREEN (CONTINUED)
C. Touch Screen Calibration
Ensure patient is not present on the bed before performing the Touch Screen Calibration.
1 .
From the Configuration Screen, press the Touch Screen Calibration button, Figure 28 will appear .
NOTE: If the touch screen will not respond, you will need to push the HOB 30
on the footboard at the same time (this will enable you to get directly into the touch screen calibration).
CAUTION: When pushing the buttons above the fowler, brakes, or drive actuators may activate.
2 . Carefully press and briefly hold a stylus or your finger on the center of the target (cross-hair) shown on the
screen . Repeat as the target moves around the screen . There will be four different locations to press: upper
center, lower center, center left and center right .
3 . When the last target has been touched, the Figure 29 will be displayed to inform you the 'New calibration settings
have been measured' .
4 . Tap the screen to register saved data .
Figure 28: Touch Screen Calibration: Start
5 . If you do not want to save the new data, wait for 30 seconds to cancel saved data and keep the current setting .
NOTE: The TOUCH SCREEN CALIBRATION SCREEN is a validation of the touch screen's calibration. When the
user presses anywhere on the screen, the coordinates x and y are displayed on the lower part of the screen.
www.stryker.com
Maintenance Menu Guide
2151-109-006 REV A
0
+, Brake and Drive buttons
Figure 29: Touch Screen Calibration: Completed