INTOUCH® CRITICAL CARE BED

Guidance and Manufacturer's declaration - Electromagnetic Immunity

The InTouch® Critical Care Bed is suitable for use in the electromagnetic environment specified below . The
customer or the user of the InTouch® Critical Care Bed should ensure that it is used in such an environment .
Immunity Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrostatic fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, voltage
variations and short
interruptions on power
supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a .c . mains voltage prior to applications of the test level .
T
Return To Table of Contents
1-84

EMC Information

IEC 60601 Test Level
+6 kV contact
+8 kV air
+2 kV for power
supply lines
+1 kV for input/
output lines
+8 kV differential mode
+8 kV differential mode
+2 kV common mode
+2 kV common mode
<5%Ut (>95% dip in Ut)
<5%Ut (>95% dip in Ut)
for 0,5 cycle
40%Ut (60% dip in Ut)
40%Ut (60% dip in Ut)
for 5 cycles
70%Ut (30% dip in Ut)
70%Ut (30% dip in Ut)
for 25 cycles .
<5% Ut (>95% dip in Ut)
for 5 sec .
3 A/m
2151-109-006 REV A
Compliance Level
Environment Guidance
+6 kV contact
Floors
+8 kV air
concrete, or ceramic tile .
floors are covered with synthetic
material, the relative humidity
should be at least 30% .
+2 kV for power
Main power quality should be
supply lines
that of a typical commercial or
+1 kV for input/
hospital environment .
output lines
Main power quality is that
of typical commercial and/or
hospital environment .
Main power quality should be
for 0,5 cycle
that of a typical commercial
and/or hospital environment . If
for 5 cycles
the user of the InTouch® Critical
Care Bed requires continued
for 25 cycles .
operation during power main
<5% Ut (>95%
interruptions, it is recommended
dip in Ut)
that the device be powered
for 5 sec .
from an uninterrupted power
supply or a battery .
3 A/m
Power
fields should be at levels
characteristic
location in a typical commercial
and/or hospital environment .
Electromagnetic
should
be
wood,
If
frequency
magnetic
of
a
typical
www.stryker.com