Beds require an effective maintenance program, we recommend checking these items annually . Use this sheet for your
records . Keep on file .

CHECKLIST

_____ All fasteners secure (reference all assembly prints) .
_____ Engage brake pedal and push on the bed to ensure all casters lock securely .
_____ "Brake Not Set" LED on the footboard and head end siderails when brakes are not engaged .
_____ Locking steer caster engages and disengages properly (2151/2153/2155 models only) .
_____ Siderails move, latch and stow properly .
_____ CPR release working properly .
_____ I . V . pole working properly (if on bed) .
_____ No cracks or splits in head or footboards .
_____ No rips or cracks in mattress cover .
_____ All functions on head end siderails working properly (including LEDs) .
_____ All functions on footboard working properly (including LEDs) .
_____ Scale and bed exit system working properly .
_____ Night light working properly .
_____ Power cord not frayed .
_____ No cables worn or pinched .
_____ All electrical connections tight .
_____ All grounds secure to the frame .
_____ Ground impedance not more than 100 milliohms .
_____ Current leakage not more than 300 microamps .
_____ Engage drive wheel and ensure it is operating properly (Zoom
_____ Motion release switches working properly (Zoom
_____ Confirm head end Zoom
_____ Replace batteries every two years .
_____ Replace batteries immediately if batteries are observed to have corrosion at the terminals, display cracking,
have expanded or are bulging at the sides, or no longer can maintain a full charge . Use only QDF9188 for
battery replacement .
_____ Ensure ground chains are clean, intact, and have at least two links touching the floor .
_____ Check fowler angle for accuracy 0° - 70° .
_____ Check gatch angle for accuracy 0° - 15° .
_____ Check foot section angle for accuracy 0° - 50° .
_____ Siderail switches working properly (iBed Awareness option) .
_____ iBed Awareness Light Bar LEDs on footboard and siderails working properly (iBed Awareness option) .
_____ Inspect footboard control labeling for signs of degradation .
_____ Ensure calibration of the touch screen is accurate .
_____ Ensure calibration of the bed is accurate .
Bed Serial Number:
Completed by: ___________________________________________
www.stryker.com

Preventative Maintenance

option - 2152/2154/2156 model only) .
®
®
handle functionality (2152/2154/2156 model only) .
________________________
________________________
________________________
________________________
2151-109-006 REV A
®
option - 2152/2154/2156 model only) .
________________________
________________________
________________________
________________________
Date: _________________