Problem / Failure
No Trendelenburg Motion .
No Reverse Trendelenburg Motion .
No Cardiac Chair Motion .
HOB 30
0
No Vascular Motion
www.stryker.com
Troubleshooting Guide
1 .
Check the touch screen trend angle display for accuracy of the level
of the litter .
I .
If not accurate, recalibrate the bed (refer to the Bed
Calibration procedures located on
a . If recalibration does not work, replace the trend angle
sensor and recalibrate .
b . If replacement of the trend angle sensor did not resolve
the problem, replace the CPU/Power board .
1 .
Check the touch screen trend angle display for accuracy for the
accuracy of the level of the litter .
I .
If not accurate, recalibrate the bed (refer to the Bed
Calibration procedures located on
a . If recalibration does not work, replace the trend angle
sensor and recalibrate .
b . If replacement of the trend angle sensor did not resolve
the problem, replace the CPU/Power board .
1 .
Check the touch screen fowler and foot section angle display for
accuracy for the accuracy of the angle of the fowler and foot section .
I .
If not accurate, recalibrate the bed (refer to the Bed
Calibration procedures located on
a . If recalibration does not work, replace the fowler or foot
section angle sensor depending on the one which was
not accurate, then recalibrate .
b . If replacement of the fowler or foot section angle
sensor did not resolve the problem, replace the CPU/
Power board .
1 .
Check the touch screen fowler angle display for accuracy of the
angle of the fowler .
I .
If not accurate, recalibrate the bed (refer to the Bed
Calibration procedures located on
a . If recalibration does not work, replace the fowler angle
sensor and recalibrate .
b . If replacement of the fowler angle sensor did not
resolve the problem, replace the CPU/Power board .
1 .
Check the touch screen fowler, gatch, foot, and trend/rev . trend
display for accuracy of the angle of all .
I .
If not accurate, recalibrate the bed (refer to the Bed
Calibration procedures located on
a . If recalibration does not work, replace the angle
sensor of the section that is not accurate and then
recalibrate the bed (refer to the Bed Calibration
procedures located on
b . If replacement of the fowler angle sensor did not
resolve the problem, replace the CPU/Power board .
2151-109-006 REV A
Recommended Action
page
1-17) .
page
1-17) .
page
1-17) .
page
1-17) .
page
1-17) .
page
1-17) .
Return To Table of Contents