WARNING (CONTINUED)
• The Bed Exit system is intended only to aid in the detection of a patient exiting the bed . It is not intended to replace
patient monitoring protocol . The Bed Exit system signals when a patient is about to exit the bed .
• The Bed Exit system is not designed to be used with patients weighing less than 50 lbs (23 kg) .
• Do not steam clean, hose off or ultrasonically clean the bed . Do not immerse any part of the bed . The internal
electrical parts may be damaged by exposure to water . Hand wash regularly all surfaces of the bed with warm water
and a mild detergent . Wipe cleaned surfaces dry to avoid build-up of cleaning substance . Inspect the mattress
after each use . Discontinue use if any cracks or rips, which may allow fluid to enter the mattress, are found in the
mattress cover . Failure to properly clean the mattress, or dispose of it if defective, may increase the risk of exposure
to pathogenic substances and may bring about diseases to the patient and user .
• Always unplug the bed power cord from the wall socket when servicing or cleaning the bed . When working under the
bed with the bed in the high position, always apply the brakes and place blocks under the Bed Lift levers to prevent
injury in case the Bed Down switch is accidentally pressed .
CAUTION
• Preventative maintenance should be performed at least once a year to ensure all bed features are functioning
properly . Ensure that any bed malfunction is promptly reported to your service personnel for immediate attention .
• When large fluid spills occur in the area of the circuit boards, cables and motors, immediately unplug the bed power
cord from the wall socket, remove the patient from the bed and clean up the fluid . Have maintenance completely
check the bed . Fluids can have an adverse effect on operational capabilities of any electrical product . DO NOT
put the bed back into service until it is completely dried and has been thoroughly tested for safe operation . Ensure,
among other things, that the plastic components being used as covers for the siderail mechanism arms and the foot
end casing are removed and that the parts they cover are thoroughly dried .
• To avoid damage to the siderail mechanisms, do not move the bed using the raised siderails . Move the bed using
the handles integrated to the boards .
• Because individual beds may have different options, footboards should not be moved from one bed to another .
Mixing footboards could result in unpredictable bed operation .
• Leave the frame in the lowest position when the patient is unattended to help reduce the number and severity of a
potential fall .
• The mattress thickness should not exceed 6 inches (15 .24 cm) .
• When using a mattress thicker than 6 inches or mattress overlay, extra caution and/or patient supervision may be
required to reduce the likelihood of occurrence of a patient fall .
• When servicing use only identical replacement parts provided by Stryker .
• For the batteries to maintain sufficient charge, you must plug the bed into a wall outlet (regulated AC power source)
when not in use to maximize bed performance while on battery power .
• If batteries are observed to have corrosion at the terminals, display cracking, have expanded or are bulging at the
sides, or no longer can maintain a full charge, they should be replaced immediately .
• When replacing the batteries, use only Stryker authorized batteries . Use of non-Stryker batteries may lead to
unpredictable system performance .
• Upon a Battery Low alarm (Battery Low LED on Footboard and/or audible beep), stop using the Zoom
recharge the batteries immediately . Ignoring the Battery Low alarms may cause your batteries to degrade quicker
than normal and may decrease battery life .
Note
• Throughout this Maintenance Manual, the words "right" and "left" refer to the right and left sides of a patient lying
on his/her back on the bed .
• The addition of accessories affects the motion of the bed .
• The iBed option is only used as an informational feature and is not supposed to replace normal patient monitoring
protocol .
Return To Table of Contents
1-10
Summary of Safety Precautions
2151-109-006 REV A
®