Herunterladen Inhalt Inhalt Diese Seite drucken

Filling The Pump - Medtronic SYNCHROMED EL serie Gebrauchsanweisung

Programmierbare pumpen
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

PHM036En2.qxd
2/16/00 12:03 PM
Page 26
26
Instructions for Use
35 °C - 40 °C
Figure 3. Empty pump while it is submerged in warm bath.
CAUTION:
Do not allow air into the pump's reservoir through an open
needle (Figure 4). This can result in reservoir contamination or
activation of the reservoir valve.
Figure 4. Do not expose the pump reservoir to air
3.
When all fluid has been removed from the pump's reservoir, remove needle
and syringe containing sterile water, while maintaining negative pressure on
the syringe.

Filling the Pump

Do not fill a pump that is not adequately warmed. Do not
CAUTION:
allow the pump to cool below 35 °C (95 °F) during the initial
fill procedure. To maintain temperature and proper reservoir
pressure of pump, fill the drug reservoir while the pump is
submerged in the sterile warm water bath.
Medtronic - Kerkrade / PARTNUMBER :
through an open needle.
198076-036
SHEETNR. :

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis