20
Pump Operations
Parameter
Drug Name
Pump Model
Metrology and Concentration
Calibration Constant
Infusion Mode
Rate
Reservoir Volume
Low Reservoir Alarm Status
Low Battery Alarm Status
Date of Change
1
Indicates 4 digit pump model number; pump model determines presence of side catheter access port and
absence of reservoir valve.
2
Indicates a 2-3 digit number.
3
Approximately 0.3-0.5 microliters/hour.
4
Approximately 5 to 9 ml or 13 to 17 ml, depending on reservoir volume. Initial interrogation of the pump
at implant may display a lower volume. Actual volume will be within the 5 to 9 ml or 13 to 17 ml range
depending on reservoir volume. Reservoir volume is an electronic prediction of what should be in the
reservoir. When the pump is in its sterile package, the port is sealed and this prevents any fluid flow
from the pump.
Medtronic - Kerkrade / PARTNUMBER :
Table 4. Factory-set information.
Preset Data
Water
1
NNNN
-10 or NNNN
1000.0 µl/ml
2
NNN
Simple Continuous
3
N.N
4
NN.N
Enabled
Disabled
Date and Time
198076-036
1
-18
SHEETNR. :