GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The Compex stimulator is intended for use in the electromagnetic environment specified below. The customer or the user of the Compex
IMMUNITY TEST
Electrostatic discharge
(ESD) IEC 61000-4-2
±2kV for power supply
Electrical fast transient/
burst
IEC 61000-4-4
±1kV for input/output
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply
input lines
(30% dip in UT) for 25
IEC 61000-4-11 a
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
stimulator should assure that it is used in such an environment.
IEC 60601
COMPLIANCE LEVEL
TEST LEVEL
±6kV contact
±8kV air
±2kV (power lines)
lines
lines
±1kV Line to Line (or
Neutral)
±2kV line(s) to earth
<5% UT
(>95% dip in UT) for
(>95% dip in UT) for
0,5 cycle
40% UT
(60% dip in UT) for 5
(60% dip in UT) for 5
cycles
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(>95% dip in UT) for
5 sec
3 A/m
a UT is the a.c mains voltage prior to application of the test level.
ELECTROMAGNETIC ENVIRONMENT -
±6kV contact
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
±8kV air
relative humidity should be at least 30%.
Mains power quality should be that of a typical
Not Applicable
(I/O lines)
±1kV Line to Line
Mains power quality should be that of a typical
Not Applicable
(Line to Earth)
<5% UT
0,5 cycle
40% UT
Mains power quality should be that of a typical
commercial or hospital environment. If the user
cycles
of the Compex stimulator requires continued
operation during power mains interruptions, it
70% UT
is recommended that the Compex stimulator be
powered from an uninterrupted power supply or
cycles
<5% UT
(>95% dip in UT)
for 5 sec
Power frequency magnetic fields should be at
3 A/m
levels characteristic of a typical location in a typical
20
GUIDANCE
commercial or hospital environment.
commercial or hospital environment.
a battery.
commercial or hospital environment.