en
Mechanical alkaline cleaning and thermal disinfection
Type of device: Single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Neutralization
IV
Intermediate rinse I
V
Intermediate rinse II
VI
Thermal disinfection
VII
Drying
D–W:
Drinking water
DI–W:
Deionized water
1)
The cleaning and disinfection devices listed below were used to verify cleanability: Miele 7836 CD
2)
Prion-deactivating detergent (see Specifications Dr. Weigert neodisher® SeptoClean)
►
Place the product in a tray that is suitable for
cleaning (avoiding rinsing blind spots)
►
Open the links and joints before placing the product
on the sterilization tray.
►
Connect components with lumens and channels
directly to the rinsing port of the injector carriage.
►
Check visible surfaces for residues after mechanical
cleaning/disinfecting.
16
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD–W
>10/50
2
DI–W
>10/50
1
FD–W
>10/50
1
DI–W
90/194
5
FD–W
-
-
-
9.9
1)
Chemicals
-
Dr. Weigert neodisher® SeptoClean 1%
Working solution
B. Braun Helimatic® Neutralizer C
Working solution 0.15%
-
-
-
In accordance with the program for the
cleaning and disinfecting machine
Inspection, maintenance and
checks
Damage (metal seizure/friction
corrosion) to the product
caused by insufficient lubrica-
CAUTION
tion!
►
Prior to function checks,
lubricate moving parts (e.g.
joints, pusher components
and threaded rods) with
maintenance oil suitable for
the respective sterilization
process (e.g. for steam ster-
ilization: Aesculap
STERILIT® I oil spray JG600
or STERILIT® I drip lubrica-
tor JG598).
2)