9.8
Mechanical cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must in principle
have a tested effectiveness (e.g. compliance with
EN ISO 15883).
Note
The cleaning and disinfection machine used for pro-
cessing must be serviced and checked at regular inter-
vals.
Manual pre-cleaning using brush
Phase
Step
I
Cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
Phase I
►
Clean the product using a suitable cleaning brush
in the solution until there is no residue visible on
the surface.
►
If applicable, use a suitable cleaning brush to brush
any surfaces that are not visible for at least 1 min.
►
Mobilize non-rigid components, such as set screws
and hinges, during cleaning.
►
Then flush these areas thoroughly at least five
times with the cleaning disinfectant solution using
a disposable syringe (20 ml).
Phase II
►
Rinse/flush the instrument thoroughly (all accessi-
ble surfaces) under running water.
►
Mobilize non-rigid components, such as set screws
and hinges, during rinsing.
T
t
Conc.
[°C/°F]
[min]
[%]
RT
>15
1
(cold)
RT
1
-
(cold)
Water
Chemicals
quality
D–W
B. Braun Stabimed® fresh
D–W
-
en
15