Guidance and manufacturer's declaration – electromagnetic immunity –
AUTOCON
®
ll 200 model 20 5322 20 is intended for use in the electromagnetic environment speci ed below. The customer or user of the
AUTOCON
EN/IEC 60601
Immunity test
test level
Conducted RF
3 V rms
IEC 61000-4-6
150 kHz to < 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and re ection by structures, objects and
people.
a Field strengths from xed transmitters, such as base stations for radio [cellular/cordless] telephones and land mobile radios, amateur radio, AM
and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed
RF transmitters, an electromagnetic site survey should be considered. If the measured eld strength in the location in which the unit is used exceeds
the applicable RF compliance level above, the unit should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the unit.
b Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
Appendix
Electromagnetic Compatibility
(EMC) Information
Table 204
for equipment and systems that are not life-supporting
®
|ll|200 model 20 5322 20 should ensure that it is used in such an environment.
Compliance level
3 V rms
3 V/m
(
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the AUTOCON
20 5322 20, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency
of the transmitter.
Recommended separation distance:
3
5 .
3 /
150 kHz to < 80 MHz
3
5 .
3 /
80 MHz to < 800 MHz
7
3 /
800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in
watts [W] according to the transmitter manufacturer and d is the
recommended separation in meters [m].
Field strengths from xed RF transmitters, as determined by an
a
electromagnetic site survey
, should be less than the compliance
b
level in each frequency range
.
Interference may occur in the vicinity of equipment marked with the