3
WARNING: The mode and power set-
ting should be adapted to the interventi-
on and should be as low as possible.
3
WARNING: When performing unipo-
lar HF surgery on bio-tissue with ne
structures, there is the risk of burns due
to local current concentration. For this
reason, bipolar HF surgery is recom-
mended in these areas.
47
Begriffsbestimmungen
und Erläuterungen zum
AUTOCON
®
II 200
Factory settings for HF modes
Basic program
Standard settings for free program memory slots
UNIPOLAR
POWER Cut
STANDARD Coag
FORCED Coag
GASTRO CUT (optional)
BIPOLAR
STANDARD Coag with AUTO STOP
STANDARD Coag
LAP
Cut mode
Bipolar
Unipolar
POWER Cut
GYN HYS
Cut mode
Bipolar
Unipolar
POWER Cut
GYN LASH
Cut mode
Bipolar
Unipolar
POWER Cut
BIPOL AUTO
Cut mode
Bipolar
Unipolar
POWER Cut
Gastro
Cut mode
Bipolar
Unipolar
GASTRO Cut
Definitions and explanations
for AUTOCON
®
II 200
Effect
Output [W]
3
150
3
40
2
60
3
120
Effect
Output [W]
3
35
3
35
®
Display – AUTOCON
Cut effect
Cut performance
3
35
Display – AUTOCON
®
Cut effect
Cut performance
4
160
Display – AUTOCON
®
Cut effect
Cut performance
4
160
®
Display – AUTOCON
Cut effect
Cut performance
4
45
®
Display – AUTOCON
Cut effect
Cut performance
3
120
AUTOCON
Time
Interval
2
8
Time
Interval
II 200 following programming
Coag mode
Coag effect
Coag performance Type of activation
Standard Coag
4
Forced Coag
3
II 200 following programming
Coag mode
Coag effect
Coag performance Type of activation
Forced Coag
3
II 200 following programming
Coag mode
Coag effect
Coag performance Type of activation
Standard Coag
4
Coag off
II 200 following programming
Coag mode
Coag effect
Coag performance Type of activation
Standard Coag
4
Forced Coag
3
II 200 following programming
Coag mode
Coag effect
Coag performance Type of activation
Forced Coag
2
II 200
®
Type of activation
Off
Off
Off
Off
Off, Auto Start deactivated
Off, Auto Start deactivated
35
Single-pedal
40
Dual-pedal
50
Dual-pedal
35
Dual-pedal blue
Dual-pedal yellow
35
AUTO START
40
Dual-pedal
50
Dual-pedal

Werbung