The emptying time starts immediately after checkweighing. Emptying is
regarded as complete when the emptying time is exceeded, regardless of
whether the scale is empty or not. During the emptying time a control output
can optionally be activated. When emptying is complete, the end of the filling
process is indicated by the Finished signal.
Information
Emptying starts immediately after checkweighing. Emptying is regarded as
complete when the gross weight is below the empty weight or when the
emptying time is exceeded, regardless of whether the scale is empty or not.
When emptying is complete, the end of the filling process is indicated by the
Finished signal.
9.2.9

Optimization

When Optimization is active (parameter P1 > 0), the coarse flow and fine flow
phases are optimized by the sensor electronics. The minimum fine flow weight
(FFM) ensures compliance with the tolerances. This limit prevents the fine flow
component being minimized further by increasing the coarse flow component.
Menu
Optimization
WTX120
Key
Function/secondary selection
0
Optimization deactivated
1
Highest degree of optimization, correction
factor 1
2
Medium degree of optimization, correction
factor 0.5
3
Lowest degree of optimization, correction
factor 0.25