Herunterladen Diese Seite drucken

Adaptation To The Weighing Environment; Calibration And Geo Value Setting - HBM WTX120 Bedienungsanleitung

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

10.2.1 Adaptation to the weighing environment

To obtain an optimal weighing result, the following parameters can be set to
adapt the A/D converter to the weighing environment:
Information
Operation subject to weights and measures legislation is available in Standard
mode / Non-automatic weighing instrument (NSW)
Menu
Motion window
Motion counter
Filter size
Auto zero range
Pushbutton zero
Power-up Zero
Overload
NTEP approval?
Underload 20d?
Update rate

10.2.2 Calibration and geo value setting

Before calibrating, the calibration location must be set based on the geo value
table (e.g. Germany = 20). This allows a scale to be precalibrated at the
factory. When subsequently installing the scale at a location with differing
gravitational pull, all that is needed is to enter the geo value of the installation
location. The scale does not have to be recalibrated in such a case.
WTX120
Function
Size of the motion window for standstill monitoring
Number of weight values (movement counter)
Filtration level of the digital filter (with an unstabilized weight
indicator)
Size of range for automatic zeroing
Size of the range of zero setting for Zero key
Size of the power-up range of zero setting
Weight limit for hiding overload on the display
Metrological properties according to American requirements
Hide weight indicator if gross weight values less than -20d
Select the update rate for the digital filter
A04500_02_Y00_00 HBM: public
Calibration
123

Werbung

loading