- Send the "Adjust zero load" command (input word 0, bit 7) in accordance
with the signal exchange (see chapter 12.4 "Description of signal
exchange")
If successful, you will receive a confirmation bit and the zero point is set.
2. Calibration of the maximum capacity
- Set the "maximum capacity" (input words 50/51) to 1,000,000 (= 2mV/V)
- Send the "Adjust maximum capacity" command (input word 0, bit 8) in
accordance with the signal exchange (see chapter 12.4 "Description of
signal exchange")
If the calibration is successful, the status bit (input word 5, bit 15) is set to 1.