Calibration
10.3 Group selection (group 1-9)
For more information on the keys required to access this menu group, go to
the
Navigation
Pilot.
Menu
Select group 1-9
Save parameters?
→ If switch S1 up/saved
Error Calibr. Jumper
126
Key
Function
Select the parameter groups
Scroll between groups 1 - 9
1 Scale parametersSet weighing parameters
2 Calibration
3 Linearization
4 Zero adjust
5 Adaptation
6 High resolution Display with 10 times
7 Reset
8 Calculate span
9 W&M info
Save and return to main menu
Finalize selection
To save calibration data:
Y
Y
N
Warning: Switch S1 not in calibration position,
parameters cannot be saved!
A04500_02_Y00_00 HBM: public
Calibrate scale
Linearize scale
Adjust zero range
Adapt to the environ
ment
resolution
Reset parameters
Calibrating without
weights
Check the approval
parameters
Save data to EEPROM
Discard all changes
WTX120