Herunterladen Diese Seite drucken

HBM WTX120 Bedienungsanleitung Seite 171

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Notice
Danger of short circuit The battery's positive terminal must not come into
contact with the housing when replacing the battery!
Notice
On legally verified calibrated systems, access to the battery is protected with a
verification seal. If the housing panel is opened the seal will be damaged!
Notice
Replace the batteries of legally verified calibrated systems every time a new
calibration is carried out!
Notice
Insert a new battery within 30 seconds, otherwise the stored data will be lost!
The lithium battery backing up the RAM and the real-time clock will last at least
three years during normal operation. The battery must be replaced by a trained
service technician every three years.
Menu
Setup/limit switches
(standard application)
WTX120
Transport, maintenance and cleaning
Values
LS1 source: OFF
LS2 source: OFF
A04500_02_Y00_00 HBM: public
Values
LS3 source: OFF
LS4 source: OFF
169

Werbung

loading