Calibration
Menu
Overload
NTEP approval?
Underload 20d?
With Taring?
140
Key
Function/secondary selection
9999
N
Y
Y
A04500_02_Y00_00 HBM: public
Overload display suppression (in D) can be
set here. If the scale exceeds the value
(maximum load + overload), the display is
hidden.
Example: Overload 9
On multi-range weighing machine, 1d
corresponds to the digit of the lowest weighing
range.
Switch to the metrological properties according
to American requirements (National Type
Evaluation Program) as per Handbook 44.
N
OFF (default)
ON
Y
Must be set to Y(es) for applications in the
USA, Canada and Australia!
Hide weight indicator if gross weight values
less than -20d (below zero)
OFF
N
ON
Y
For applications in accordance with OIML R76,
Underload 20d must be set to Y
If activated, the analog output is set to 0mA if
the value falls below -20d.
Select whether tare functions are disabled or
enabled.
N
Tara functions disabled
Y
Tara functions enabled (default)
For scales that only operate in Gross mode,
the tare functions can be disabled here so as
to skip cost-intensive checking of the tared
scale during legal verification.