Calibration
When a tared scale (net weighing) is set to zero, no automatic switch is made
to the lower digit, and the tare is not deleted. To switch back to the low digit,
the tare must first be deleted with the Tare key. Automatic switching is only
carried out on a scale that has not been tared (gross weighing).
Example of a weight characteristic over time and the displayed scale interval:
Gross weight/kg
Scale interval/kg
Fig. 11.1
Example of a multi-range weighing machine
Tare balancing and tare input values are carried over into the higher range
when load is applied, and are automatically rounded to the higher scale
interval.
The multi-range weighing machine permits configuration of a single SG base
unit for both low and high loads.
When configuring the multi-range weighing machine, all ranges and scale
intervals can be freely selected; only the number of decimal places has to be
the same for all ranges. If 0.5 kg and 1 kg scale intervals are used, for
example, 0.5 and 1.0 must be entered as the intervals. Additionally, the
maximum resolution of the load cells being used (e.g. 3000d) must not be
exceeded in any of the set ranges.
WTX120
121
A04500_02_Y00_00 HBM: public