Calibration
→ If the internal resolution is too low, the following warning is displayed
Resolution error
Menu
Select Group
Save Parameters?
Information
To discard the calibration values, you must exit Setup without saving (Save
Parameters No) before you can call up another group.

10.6 Linearization

For more information on the keys required to access this menu group, go to
the
Navigation
Pilot.
Linearization compensates for any variation of the scale characteristic (from
the ideal straight line) by setting up to 6 correction points adapted to the actual
scale characteristic.
The linearization software supports up to 6 correction points, though in practice
1 or 2 are almost always enough. The linearization points should be placed at
the points on the weight curve where the greatest variations occur.
Linearization points must be lower than the set maximum load. The chapters
between adjacent linearization points are calculated as straight, meaning the
variations are interpolated in linear mode.
134
The internal resolution should be at least 10
times the set resolution.
Check calibration data and repeat calibration if
necessary.
Key
Function/secondary selection
Y
N
The calibration values are saved when you exit
the Select Group step, provided you answer
the Save Parameters? prompt with Y
A04500_02_Y00_00 HBM: public
Parameters are saved
Parameters are not saved
WTX120