Herunterladen Diese Seite drucken

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Mastermode
Menu
3 Mastermode
Mastermode:
Damping
Type
→ Only with Critical Damped, Butterworth, Bessel
Passes
8.1

Filter settings

The WTX supports a total of four filter modes (type): Critically damped,
Butterworth, Bessel and Standard. Standard is identical to Critically
damped with a fixed setting of Passes=4. Passes between 0 and 20 can be
selected for the first three filters. A Filter size of between 0 and 20 can be
selected for all of them. The Type and Passes can be set in the
Mastermode->Adapt menu (Passes only if Standard is not selected as the
Type).The Filter size is set in the
Service Mode->Calibration->Adaption menu.
90
Key
Adapt
0
Standard
2
A04500_02_Y00_00 HBM: public
Function/secondary selection
Damping
Here the Filter Size selected in calibration
mode can be reduced or increased by up to
5 increments. The default setting is 0.
-5 to +5
Adapt the filtration level
Note: Also possible when switch S1 is acti­
vated.
The characteristic of the digital filter can be set
here:
Standard
Critical Damped
Butterworth
Bessel
Set the filter order
1-20
WTX120

Werbung