Herunterladen Diese Seite drucken

Adaptation; Navigation - HBM WTX120 Bedienungsanleitung

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Calibration

10.8 Adaptation

For more information on the keys required to access this menu group, go to
the

Navigation

Pilot.
Menu
Select group 1-9
Motion Window
MotionCounter
Filter Size
138
Key
0.5D
99
OFF
A04500_02_Y00_00 HBM: public
Function/secondary selection
Press 5x for group 5 Adaptation
Motion window
Here the window is set in digits (increments),
which must contain consecutive weight values
for the A/D converter to signal a stationary
weight value.
OFF
3.0D, 2.0D, 1.0D, 0.5D
Time span for standstill recognition
Here the time span is set in which consecutive
weight values must occur in the motion
window for standstill to be detected.
1 unit = 0.02 sec.
Example: 20 = 0.4 sec
Filtration level
The filtration level of the digital filter can be set
here. The following settings are possible:
Off
0
1 to 20
Light to heavy filtering.
With a very unstable scale display (such as in
the case of live cattle weighers) heavy filtering
is recommended.
Motion control off
Window size
WTX120

Werbung