Error message display
Power-up
Out of Range
Power-up motion
Transfer error
WTX120
Possible causes
Over/under power-up zero
setting range. This
message appears
immediately after activation
if the scale is loaded with a
weight greater than the set
power-up zero range
(+2%, +10%) or less than
the set power-up zero
range (-2%, -10%).
Power-up motion.
This message appears
immediately following
activation if the scale does
not find a stable weight
value within the set
power-up zero range
(±2%, ±10%).
Computer off or not ready
Transfer cable defective or
plug not attached
A04500_02_Y00_00 HBM: public
Remedy
Unload scale or
load scale as appropriate
Stabilize the scale
Turn on computer or start
receive program
Check transfer cable
and plug
In an emergency, disable
data transfer in entries
Faults
177