4.9.1

Status LEDs

D600 (red)
ON
Flashing
OFF
Tab. 4.5
D600 system failure (red/green)
D601 (red)
ON
Flashing
OFF
Tab. 4.6
D601 bus failure (red/green)
4.9.2

Configuring PROFINET variants

The weighing terminal works as a PROFINET I/O device, forming an I/O inter­
face with up to 64 input words and up to 64 output words, and is suitable for
transfer speeds up to 100 MBit/s.
The IP address of the PROFINET module is set on the PLC (e.g. S7, TIA) and
not on the weighing terminal. The IP address can also be set using the
1
Hilscher tool
)
.
To set the WTX120 parameters in the PROFINET variants, a GSDML file is
required which you can download from
trielles-eichfaehiges-waegeterminal/
Information
The IP address for PROFINET is not the same IP address as for the Ethernet
TCP/IP connection on the underside of the device.
1)
Hilscher Ethernet Devices Configuration
net+Device+Configuration
WTX120
SF-LED (system failure)
Time out / diagnosis available / system error
DCP signal service is initiated via the bus
No error
BF-LED (bus failure)
No configuration; low Ethernet speed; no Ethernet link
No data is exchanged
No error
in the WTX120 device category.