Calibration
10
Calibration
To access the Calibration group select the following keys:
Menu
Standard
WTX120 V#.#
Password
1 Service mode
Service:
Interface
Service:
Calibration
10.1 Multi-range
The A/D converter can be configured as a single, dual or triple-range scale.
The individual ranges differ in their maximum load and scale interval.
Example of three-range scale with a resolution of 3000 increments in each
weighing range:
Lowest weighing range: 0 to 1500kg / 0.5 kg
Middle weighing range: 0 to 3000kg / 1.0 kg
Highest weighing range: 0 to 6000kg / 2.0 kg
The multi-range weighing machine automatically switches from the lower range
to the next higher range when under load.
120
Key
Function
Switch from Standard to version notice
Display the current firmware version number
Call up password entry within three seconds
Enter the password
324
Call up Service Mode
Press twice to switch to the Calibration
group option