Approved Applications - HBM WTX120 Bedienungsanleitung

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Calibration

10.2.3 Approved applications

The maximum resolution (maximum load/interval) of approved scales is 10,000
increments. Additionally, the following settings must be implemented for the
parameters in the Adaptation group:
Information
Operation subject to weights and measures legislation is available in Standard
mode / Non-automatic weighing instrument (NSW)
Menu
Motion window
Motion counter
Auto zero range
Pushbutton zero
Power-up Zero
Overload
1)
Subject to certain preconditions, the OFF setting is also allowable here, such as for tank weighers.
After calibrating the scale, the approval parameters can be saved to the
EEPROM of the A/D converter, where they are retained even in the event of a
power outage. In approved applications, the covering and lid must be sealed
with adhesive labels so that the position of the calibration switch can no longer
be changed.
► Unscrew the covering from the calibration certificate on the front of the
housing. You can use the small calibration switch S1 that is now visible to
save the calibration data to the EEPROM that is protected against power
outages:
124
Function
Size of the motion window for standstill
monitoring
Filtration level of the digital filter (with an
unstabilized weight indicator)
Size of range for automatic zeroing
Size of the range of zero setting for Zero key
Size of the power-up range of zero setting
Weight limit for hiding overload on the display
A04500_02_Y00_00 HBM: public
0.5D
≥7
0.5D
±2 %
±2% or
1)
±10%
max. 9 D
WTX120