To save the zero balance before calibrating the weighing range, you must exit
Setup at this point. Answer the Save Parameters question with Y. Then you
can call group 2 again to resume the calibration, skipping the zero calibration.
Menu
Calibrate Load?
Calibr.Weight
Calibrating...
Load:
Load(mV/V):
Information
The Clr key can be used to clear the value and enter a new one.
WTX120
Key
Y
999999
9999999
1.40000
A04500_02_Y00_00 HBM: public
Function/secondary selection
Calibrate weighing range
Calibrate range
Y
N
Continue to step Load(mV/V)
Note: It is recommended that you use as large
a calibration weight as possible.
Suggestion for calibration weight
(=weighing range)
Clear the suggested weight value and enter
any partial load value
Mount load and confirm; continue with
calibration
Measurement of the load signal
Display the 10 times higher weight value
resolution (as a check)
Display the standardized signal - see
Calculate Span
Note: The signal can be calculated in millivolts
from the displayed mV/V value by multiplying it
by the Excitation value (5V or 10V).
Calibration