Herunterladen Diese Seite drucken

HBM WTX120 Bedienungsanleitung Seite 127

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

S1
Top switch:
Calibration data saved, changes not possible
Bottom switch:
(Direction of connections)
Possible to change the calibration data
Fig. 11.3
Calibration switch S1 WTX120
Information
Position of the S1 switch (under the covering)
In approved applications, the covering and lid must be sealed with adhesive
labels so that the position of the calibration switch can no longer be changed.
→ If switch S1 up/saved
Calibration locked
WTX120
S1
Switch S1 not in calibration position,
parameters cannot be saved!
Access calibration without save option
(e.g. for checking)
A04500_02_Y00_00 HBM: public
Calibration
WTX
120