Onscreen typeplate? Y
Update Rate

10.9 High resolution

For more information on the keys required to access this menu group, go to
the
Navigation
Pilot.
With this menu group you can display the weight value in 10 times higher
resolution.
Menu
Select group 1-9
Weight:
WTX120
Key
225
Key
9999999
A04500_02_Y00_00 HBM: public
Function/secondary selection
Select whether the Onscreen typeplate
will be displayed above the scale display:
Display typeplate
Y
Hide; the field is left blank
N
Select the update rate for the digital filter
50 ... 800 Hz Setting range
Default setting
225 Hz
If there are vibrations around the scale, the
properties of the digital filter can be optimized
here. The update rate should be selected as
far away as possible from the vibration
frequency and from the multiple of the
vibration frequency.
If an Update rate >400Hz is set, this can
affect the performance when using PanelX;
communication times can be extended.
Adjusting the Update rate is not
recommended unless required by the
ambient conditions.