Calibration
→ After displaying the last linearization point, or if there are no points
New Fixpoint 1?
Enter Fixpt.1
Calibrate Fixpt?
→ With Calibrate Fixpt = Y
Linearization.
Menu
Weight:
Lin.Signal1
A new linearization point can be inserted between existing points. Then all the
points are renumbered in ascending order.

10.7 Zero adjust

For more information on the keys required to access this menu group, go to
the
Navigation
Pilot.
In this menu group the zero point of the scale can be redefined. This becomes
necessary in particular when the scale has been calibrated using an auxiliary
frame. After removing the auxiliary frame, the final zero point is then defined.
The change in zero point shifts the weighing range accordingly.
136
N
999999
N
Key
9999999
0.50000
A04500_02_Y00_00 HBM: public
Y
Enter linearization point
Do not enter linearization point; back to
N
Select Group
Enter the weight to be linearized
Calibrate linearization point
Y
N
Enter linearization value
Measurement of the linearization signal
Function/secondary selection
Display the linearized weight value with 10
times higher resolution (as a check)
Enter a calculated linearization signal or import
the values from another A/D converter
Continue in New Fixpoint 2? step