9.2.14 Target weight

This is the target weight of the batching process. Entering the target weight
automatically implements the following settings, if they were not already set.
Command
Meaning
CFD
Coarse flow cut-off point
CBK
Coarse flow fill flow monitoring
FFD
Fine flow cut-off point
FBK
Fine flow fill flow monitoring
FFM
Minimum fill flow
LTL
Lower tolerance limit
UTL
Upper tolerance limit
SYD
Systematic difference

9.2.15 Empty weight

For dynamic processes such as batching, you should limit the possible tare
value to prevent overfilling. This might happen if there is already material in the
container on starting. If you do not set this tare limit, the coarse flow cut-off
point will be used as the limit.
After starting the filling process (RUN command or digital input or Enter key),
any preset tare delay is left to elapse and a check is then completed to
determine whether the weight is below the specified value for the empty
weight. If it is, taring is carried out; otherwise no new taring is carried out (the
old tare value is retained).

9.2.16 Minimum start weight

When this value is specified, the batching process is not started if the current
gross value is less than this minimum start weight. Use this command and the
empty weight to define the range in which dosing is to start.
WTX120
A04500_02_Y00_00 HBM: public
Supervisor mode
Setting
50.0% of the target weight
0 (= switched off)
95.0% of the target weight
0 (= switched off)
1.0% of the target weight
99.8% of the target weight
100.2% of the target weight
0 (= switched off)
105