Herunterladen Diese Seite drucken

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

→ Single-range balance Two Intervals T+, dual-range balance Dual Range or
multi-interval weighing machine Two Intervals
High Capacity
High Int.
Low Capacity
→ Single-range balance Three Intervals T+, triple-range balance Triple Range
or multi-interval weighing machine Three Intervals
High Capacity
High Int.
Mid Capacity
Mid Int.
WTX120
999999
999.9999
999999
999999
999.9999
999999
999.9999
A04500_02_Y00_00 HBM: public
Enter the maximum load of the higher range.
Example: High Capacity 6000
Enter the interval of the higher range.
Example: High Int. 1.0000
The error message Entry Not Valid! is
displayed if you enter an invalid scale interval.
Enter the maximum load of the lower range.
Example: Low Capacity 3000
If the value Low Capacity is greater than
High Capacity, the error message Entry Not
Valid! is displayed.
Enter the maximum load of the highest range.
Example: High Capacity 6000
Enter the scale interval of the highest range.
Example: High Int. 1.0000
The error message Entry Not Valid! is
displayed if you enter an invalid scale interval.
Enter the maximum load of the middle range.
Example: Mid capacity 3000
If the value Mid Capacity is greater than
High Capacity, the error message
Entry Not Valid! is displayed.
Enter the scale interval of the middle range.
Example: Mid Int. 1.0000
The error message Entry Not Valid! is
displayed if you enter an invalid scale interval.

Calibration

129

Werbung

loading