HBM WTX120 Bedienungsanleitung Seite 134

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Calibration
Menu
Select group 1-9
→ Linearization points present
Fixpoints active!
Menu
Geo Value
Calibrate Zero?
Calibrating...
Zero:
Zero(mV/V):
Information
The Clr key can be used to clear the value and enter a new one.
132
Key
Key
99
Y
0.00
0.40000
A04500_02_Y00_00 HBM: public
Function/secondary selection
Press 2x for group 2 Calibration
Linearization entries have already been made;
they may influence the calibration.
Function/secondary selection
Enter the geo value. Values between 0 and 31
can be set, according to the geo value table.
The geo value of the calibration location is set.
If different, the geo value of the installation
location can be entered subsequently.
Zero balance of the scale: Unload scale
Calibrate zero point
Y
N
Continue to step Zero(mV/V)
Measure the zero point of the scale
Display the 10 times higher weight value
resolution (as a check)
Display the standardized signal - see
Calculate Span.
Note: The signal can be calculated in millivolts
from the displayed mV/V value by multiplying it
by the Excitation value (5V or 10V).
WTX120