CONTROLS, OPERATION & USE
When you are ready to begin treatment, hold the PBM adapter in contact with the target treatment area. The PBM adapter is designed to be held in a constant location for the duration of the treatment. If the
desired treatment area is larger than the PBM adapter's spot size, move the adapter to a new location and start a new treatment only after the initial treatment time has elapsed.
PBM ADVERSE EVENTS & CONTRAINDICATIONS
If patient discomfort or reddening of the skin in the treatment area occurs at any time during treatment, you can do the following:
• Defocus the laser energy by moving the adapter a few centimeters back from the skin
• Decrease the treatment time
• Stop treatment
If blistering of the skin occurs, or the patient feels a burning sensation, immediately stop treatment and rinse the area with cool water or place a cold pack on the affected area for at least 5 minutes.
Afterwards, apply a burn ointment or spray. DO NOT USE ICE.
• Do not use over articles of clothing.
• Do not treat open wounds.
• Do not apply ointment, creams, lotions, or heating lotion patches at, or in close proximity to, the treatment area.
• Do not apply therapies prior to treatment that could change body temperature, such as ultrasound, ice/heat pack, electrical stimulation, or heating patches.
• Avoid treatment sites with tattoos.
• Different implant materials will respond differently to laser energy and heat; be aware of any implants and their location; avoid direct exposure to laser energy or heat at the site of the implant.
• Excessive fatty tissue is known to transmit heat without much attenuation, therefore increase distance or decrease treatment time.
• Muscle tissue closer to the skin surface may experience a higher absorption of heat; carefully monitor skin temperature and reduce treatment time as necessary.
• Patients with swelling and/or inflammation may be sensitive to heat; reduce treatment time as necessary to ensure comfort during treatment.
• Patients with tender or sensitive skin may be hypersensitive to heat; reduce treatment time as necessary to ensure comfort during treatment.
• Scar tissue has been associated with poor circulation and reduced cooling through heat transport by blood; reduce treatment time as necessary to avoid overheating.
• Do not treat directly over the site of any known primary malignant carcinoma or secondary metastasis except for palliative care with informed consent and oncologist permission.
• Do not treat pregnant women as the effects of photobiomodulation therapy on the fetus are unknown.
PBM ADAPTER MAINTENANCE
The disposable spacers are supplied non-sterile by the manufacturer and should be wiped with isopropyl alcohol wipes by the operator prior to use. The spacers are intended for single-use only and should
never be autoclaved or reused to prevent damage or cross-contamination.
The 25 mm PBM adapter is also provided non-sterile by the manufacturer and can be wiped as needed using isopropyl alcohol wipes. Do not submerge the 25 mm PBM adapter in any type of cleaning
solution. DO NOT AUTOCLAVE the 25 mm PBM adapter.
Use the included cleaning cloth to gently wipe the glass optics of the 25 mm PBM as needed. Do not use any harsh chemicals or abrasives to clean the glass optics within the 25 mm PBM adapter. Doing so
may damage the glass.
PBM ADAPTER SPECIFICATION
Dimensions
Weight
Spot Size
Power (under preset)
Power Density
PBM DOSAGE TABLE
TIME IN SECONDS
03
06
09
12
15
18
21
24
27
30
33
36
39
42
45
48
51
54
57
60
25 MM PBM ADAPTER
7 MM PBM ADAPTER
63 x 41 x 41 mm
100 x 24.7 x 14.9 mm
76 grams
50 grams
4.91 cm
2
1.0 Watt Average
0.3 Watt Average
204 mw/cm
780 mw/cm
2
25 mm PBM Dose (J/cm
)
7 mm PBM Dose (J/cm
2
0.6
1.2
1.8
2.4
3.1
3.7
4.3
4.9
5.5
6.1
6.7
7.3
8.0
8.6
9.2
9.8
10.4
11.0
11.6
12.2
3 MM PBM ADAPTER
100 x 20.5 x 14.9 mm
50 grams
0.38 cm
0.07 cm
2
2
0.3 Watt Average
4244 mw/cm
2
)
3 mm PBM Dose (J/cm
2
2.3
12.7
4.7
25.5
7.0
38.2
9.4
50.9
11.7
63.7
14.0
76.4
16.4
89.1
18.7
101.9
21.1
114.6
23.4
127.3
25.7
140.1
28.1
152.8
30.4
165.5
32.8
178.2
35.1
191.0
37.4
203.7
39.8
216.4
42.1
229.2
44.5
241.9
46.8
254.6
2
)
2
7