WARNINGS & CAUTIONS
Failure to comply with precautions and warnings described in this User Manual may lead to exposure to dangerous optical radiation sources. Please comply with all safety instructions and warnings.
CAUTION:
• Read these instructions carefully prior to using your Gemini EVO 810+980 soft tissue diode laser.
• Ensure that all users are properly trained prior to use. Consult your distributor for training recommendations. Mandatory training on the Gemini EVO laser is done via this manual.
• Do not modify this equipment without authorization of the manufacturer.
• Laser fume and/or plume may contain viable tissue particles.
• Always wind the fiber optic cable in a clockwise manner around the fiber wrap to avoid fiber breakage.
• Do not use in the presence of combustible or combustion supporting gases.
• Always test activate the device outside the mouth before using on a patient.
• This unit has been designed and tested to meet the requirements of electromagnetic, electrostatic, and radio frequency interference standards. However, the possibility of electromagnetic
or other interference may still exist. Relocating the device may help to eliminate the interference.
• Medical electrical equipment needs special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information
provided in this manual.
• Periodically inspect laser eyewear for pitting and cracking.
• In the event abnormal performance is observed, discontinue use and follow instructions in the service and troubleshooting section of this manual.
Safety is paramount when using any energy-based surgical instrument and your office should implement a safety program for the Gemini EVO 810+980 soft tissue diode laser. If your office does
not already have a safety officer, one should be appointed to be responsible for understanding proper use, safe operation, and maintenance of the Gemini EVO laser system. Their duties should
include training office personnel in all aspects of system safety and management of the Gemini EVO laser and all accessories. / ADDITIONAL TROUBLESHOOTING: For additional troubleshooting
questions and tips call 1.801.553.4574. To check for the latest software updates, download the Gemini EVO App in the iOS or Android web store.
WARNING:
• Visible and Invisible Laser Radiation – Avoid eye or skin exposure to direct or scattered radiation. Class IV Laser Product
• Laser Safety Eye Protection MUST BE WORN by the Operator, Patient, Assistant, and anyone present when the laser is activated. Eye Protection must conform to Specification DIN EN207
Annex II of the Directive 89/686/EEC with wavelength protection of 810nm- 980nm, and ±10nm of OD 5+.
• Never direct or point the beam at a person's eyes.
• Do not look directly into the beam or at specular reflections.
• Do not aim the laser at metallic or reflective surfaces, such as surgical instruments or dental mirrors. If aimed directly at these surfaces the laser beam will reflect and create a potential
hazard.
• Never operate the laser without a fiber tip attached.
• Laser aperture at the end of the handpiece.
• Laser aperture warning label affixed to system handpiece.
• Always place the system into STBY mode when leaving the Gemini EVO 810+980 soft tissue laser unattended for a few minutes or between patients.
• Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure.
• Do not open unit housing at any time. Danger from optical radiation may exist.
• The use of accessories, other than those specified, except those supplied or sold by Ultradent Products, Inc., as replacement parts for internal or external components, may result in
increased EMMISSIONS or decreased IMMUNITY of the Gemini EVO 810+980 soft tissue laser.
WHAT IS IN THE BOX
The Gemini EVO 810+980 soft tissue laser includes the following.
Laser Unit
Activation Pedal with 2 AA Batteries
*Actual eyewear may vary from image
*Protective Eyewear (3)
3 mm, 7 mm, and 25 mm PBM
UNPACKING INSTRUCTIONS
A manufacturer or dealer representative can provide assistance when you are ready to remove the laser from its shipping container. Please do not attempt to unpack the Gemini EVO 810+980 soft tissue laser
or install the system without reading this manual first. If you are unsure about any aspect of the assembly, call your customer service representative or dealer for assistance.
SHIPPING CONTAINER INFORMATION
The shipping container you received with your Gemini EVO 810+980 soft tissue laser was specially designed to safely transport the device. In the unlikely event that you need to return the laser for service or
repair, please retain the original shipping container.
Fiber Delivery System
• Laser Warning Sign
• USB Cable
• User Manual with
Warranty Information
adapters
*Actual Power supply may vary from image
Pre-Initiated Disposable Tips (10)
*DC Power Supply with Plug Adapters
Bending tool included in the tip box
NOTE: The laser ships with the lithium ion battery and fiber delivery
system already installed
NOTE: Use proper care when transporting the unit
ALSO INCLUDED: Laser Warning Sign and User's Manual
WARNING: No modification of this equipment is allowed
EN
1