SYSTEM SPECIFICATIONS
Voice confirmation:
YES
Input: 100-240 VAC @ 50 to 60 Hz - 1.5A
Power requirements:
Output: 1.8V DC, 3.6A, 65W
Battery:
14.4V Rechargeable Lithium Ion
Bluetooth at 2.4 GHz
Wireless frequency:
Wi-Fi: 2.4 GHz / WPA2, WPA-Enterprise
Maximum Operating Altitude:
5,000 meters or 16,404 feet
The Gemini EVO 810+980 SOFT TISSUE LASER complies with the following: IEC 60825-1 / EN/S 60601-1 IEC 60601-1-2 IEC 60601-2-22 21 CFR 1040.10 and 1040.11 FCC parts 15 and 18 (47 CFR)

SERVICE AND TROUBLESHOOTING

CALIBRATION
Re-calibration is recommended every 12 months in order to assure accuracy of optical output power. The Gemini EVO 810+980 soft tissue laser may be returned to the manufacturer for re-calibration, which
you can arrange by contacting your distributor. Certain government or corporate entities may require calibration certificates which can also be provided by the manufacturer.
ADVERSE EFFECTS
If used properly, there are no known adverse effects of using the Gemini EVO 810+980 soft tissue laser. Please thoroughly read and understand all Warnings, Precautions, and Contra-indications in this
manual prior to use. In the event the laser malfunctions due to exposure to certain environment conditions, magnetic fields, external electrical influences, electrostatic discharge, pressure or variations
in pressure, accelerations, and any potential thermal ignition sources, discontinue use and follow the instructions in the service and troubleshooting section of this manual. Additional measures may be
necessary such as reorienting or relocating the device.
No separate equipment is recommended to be used to assess the favorable conditions which are acceptable for treatment or to assess the unfavorable conditions which would render a treatment
unacceptable or hazardous.
Maximum LASER OUTPUT of laser radiation with the magnitudes of cumulative measurement uncertainty and any expected increase in the measured quantities after manufacture is stated as the standard
uncertainty of measurement.
WIRELESS INTERFERENCE
This equipment has been tested and found to comply with the limits for Class B Digital Device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to
radio communications.
However, there is no guarantee that interference will not occur in a particular installation. To determine if this equipment causes harmful interference to radio or television reception, turn the equipment off
and on.
ALL OTHER CONDITIONS
In the event that the Gemini EVO 810+980 soft tissue laser fails to operate correctly, and your distributor representative is unable to help, the system will need to be returned to the manufacturer for repair.
There are no user-repairable parts available for the device. It is recommended that the system be returned in its original shipping box. If not available, one can be requested at the time you discuss your
service incident with your distributor representative.

TROUBLESHOOTING GUIDE

Why is it difficult to see the red aiming light or why is it barely visible?
CAUSE:
1. The laser is in STBY mode.
2. The aiming light under menu is turned off.
3. The fiber optic cable might be damaged or broken.
SOLUTION:
1. Aiming light is only visible under ACTIVE mode. Select ACTIVE in the Guided
Touch Interface.
2. Select menu, then aiming light. With arrows UP/DOWN adjust the aiming
light intensity.
My activation pedal is not working
CAUSE:
1. Unit is not in ACTIVE MOVE.
2. I am ACTIVE but my Activation Pedal is not working.
3. Why is my Activation pedal vibrating?
SOLUTION:
1. Make sure you select ACTIVE and tap the Activation Pedal once to wake it up.
2. Even under ACTIVE mode, you need to tap once at the pedal to wake it up.
3. Gemini EVO laser has a Haptic Sense "HS" feature that enables you to sense
small vibrations when the Activation Pedal is active. Under Menu and Haptic
Sense "HS" you can select the intensity of the vibrations or completely turn it
off.