CONTROLS, OPERATION & USE
Blue - Activation Pedal (Figure: 6.5, Option: 1)
Amber - Handpiece for PBM procedures only (Figure: 6.5, Option: 2)

09 - SOUND

To change the sound level, touch the MENU and then the SOUND icon on the Guided Touch Interface. Adjust the sound level by touching the Up / Down arrows. To exit, touch the MENU icon to save your
selection. The system remembers the last used sound setting when it is powered on. When UP arrow disappears, this indicates the volume is all the way to max and vice versa. Icon shows Red when OFF.
(Figure: 6.6)
Voice Confirmation can be enabled and disabled by touching the voice confirmation selection on the Guided Touch Interface. Red icon shows disabled and Green enabled. (Figure: 6.6)

10 - AIMING LIGHT

To change the aiming light intensity, touch the MENU and then the AIMING icon on the Guided Touch Interface. Adjust the aiming light level by touching the Up / Down arrows. To exit, touch the MENU icon to
save your selection. The system remembers the last used sound setting when it is powered on. When UP arrow disappears indicates the setting is all the way to max and vice versa. Icon shows Red when OFF.
(Figure: 6.7)

11 - PHOTOBIOMODULATION (PBM)

WARNINGS & CAUTIONS
CAUTION:
• Do not connect or disconnect a PBM adapter while the Gemini EVO laser is turned on. Only connect or disconnect a PBM adapter when the Gemini EVO laser is inactive or standby mode.
• Do not use any harsh chemicals or abrasives to clean the glass optics within a PBM adapter. Doing so may damage the glass.
• Do not autoclave the 25 mm PBM adapter or spacers. Doing so will damage the components.
• The spacers are single-use only to avoid possible cross-contamination. They must be disposed of after use in a bio hazard medical waste Sharps container.
• Wavelength-appropriate eye protection must be worn at all times while using, and in proximity of, a PBM adapter while it is in use.
WARNING:
• The PBM adapters must only be used with a Gemini EVO laser. Do not attempt to use a PBM adapter with any other laser system or light source.
• Never look directly into a PBM adapter while the laser is active, even with safety eyewear on.
• Do not use the 25 mm PBM adapter without a spacer attached.
PBM COMPONENTS
7 mm Adapter (Figure: 7.1, Option: 1) | 25 mm Spacer (Figure: 7.1, Option: 2) | 3 mm Adapter (Figure: 7.1, Option: 3) | 25 mm Adapter (Figure: 7.1, Option: 4)
PBM ASSEMBLY
(Attachment threading procedure applies to all PBM tips equally)
1. Remove the dust covers
• Dust cover / Front (Figure: 7.2, Option: 1)
• Dust cover Back (Thread) (Figure: 7.2, Option: 2)
2. Screw the PBM adapter onto the end of the Gemini EVO laser handpiece until it is tight. (Figure: 7.2, Option: 3)
3. If using the 25 mm PBM adapter, screw a spacer onto the end of the 25 mm PBM adapter. (Figure: 7.2, Option: 4)
The PBM adapter is now ready to use. To remove the PBM adapter, unscrew it from the Gemini EVO laser handpiece and re-install dust covers when not in use.
SELECTING AND ADJUSTING PBM PRESET
1. To enable Pain Relief, select PRESET on the Guided Touch Interface and navigate with the right arrow to the PAIN RELIEF category. (Figure: 7.3, Option: 1)
2. Use the UP / DOWN arrow select the desired PBM adapter. (Figure: 7.3, Option: 2)
3. Click ACTIVE to select treatment time in seconds. Display will flash 0.0 seconds. (Figure: 7.3, Option: 3)
4. Using the right arrow select the time in seconds followed by ACTIVE to enable timer. Press and hold right arrow advances timer faster in 10 seconds increments. (Figure: 7.3, Option: 4)
Laser unit is ready for PBM treatment. The timer counts down in seconds and stops automatically after treatment time is completed. If the activation pedal is released mid-treatment, the timer will pause and
will resume when the pedal is depressed again.
USE RECOMMENDATIONS
Affected muscles and/or joints have to be exposed to an adequate level of laser energy over a period of time to provide effective results. Some cases may require more than one laser treatment, or a series of
treatments, before significant improvement is reported. Repeat the treatment as necessary and monitor the progress of the patient's condition throughout the treatment.
Diode laser wavelengths, especially 810 nm, are well absorbed in melanin in the skin, which can lead to greater heating of the target tissues in patients with darker skin types. Power and treatment time
should be taken into account for patients of varying skin pigmentation. Refer to the Fitzpatrick Skin Type Scale for proper skin classification.
Pain relief preset procedure settings are programmed into the Gemini EVO laser for ease of use. Always use professional clinical judgment when selecting the laser settings for pain therapy.
Monitor the patient and adjust the power and/or treatment time as necessary to ensure both efficacy and patient comfort. The preset procedure setting is not meant to be a clinical recommendation in any
way.