ELECTROMAGNETIC ENVIRONMENT GUIDANCE
NOTES
The higher frequency range applies at 80 MHz and 800 MHz.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast, cannot be predicted
theoretically with accuracy. An investigation of the location is recommended to determine the electromagnetic environment resulting from stationary HF transmitters. If the measured field strength in the
location in which the Gemini EVO laser is used exceeds the applicable RF compliance level above, the Gemini EVO laser should be observed to verify normal operation. If unusual performance characteristics
are observed, it may be necessary to take additional measures such as reorientation or repositioning of the Gemini EVO laser.
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
WORKING CLEARANCES
The Gemini EVO laser is intended for operation in an electromagnetic environment where radiated HF interference is checked. The customer or the user of the Gemini EVO laser can help prevent
electromagnetic interference by duly observing the minimum distances between portable and/or mobile RF communication devices (transmitters) and the Gemini EVO laser. These values may vary according
to the output power of the relevant communication device as specified below.
RATED MAXIMUM OUTPUT POWER OF
TRANSMITTER [W]
0.01
0.1
1
10
100
For transmitters whose maximum nominal output is not specified in the above table, the recommended working clearance in meters (m) can be determined using the equation in the corresponding column,
where P is the maximum nominal output of the transmitter in watts (W) specified by the transmitter manufacturer.
Remark 1: The higher frequency range applies at 80 MHz and 800 MHz.
Remark 2: These guidelines may not be applicable in all cases. The propagation of electromagnetic waves is influenced by their absorption and reflection by buildings, objects, and persons.
EMERGENCY TERMINATION OF LASER EMISSIONS
The Gemini EVO 810+980 soft tissue laser has been designed with several methods to terminate emission of laser energy in emergency situations.
These methods include a power button (ON/OFF) (Figure: 21.1) and the emergency (STOP) (Figure: 21.2) button located at the front of the laser unit.

WARRANTY

Your Gemini EVO laser comes with a 2-year factory warranty. Extended warranty can be purchased and warranty certificates can be downloaded by accessing your Dashboard page at
dashboard.geminievo.com or under the warranty icon in the iOS and Android Apps.
Seller warrants the Products to be free from defects in materials and workmanship for a period of twenty-four months from the date of shipment, except for consumables. If within such period any Products
shall be proved to Seller's satisfaction to be defective, it shall be (i) repaired using new or refurbished parts, or (ii) replaced with a new or refurbished product, at Seller's sole discretion. Such repair or
replacement shall be Seller's sole obligation and Buyer's exclusive remedy under this Warranty and shall be conditioned, at Seller's option, upon return of such Products to Seller, f.o.b. its factory. This
Warranty only covers Product issues caused by defects in material or workmanship during ordinary consumer use; it does not cover Product issues caused by any other reason, including but not limited to acts
of God, modifications of or to any part of the Product, improper testing, assembly, mishandling, misuse, neglect, adjustments, alterations to the products, improper operation contrary to current instructions
relating to installation, maintenance or operation, or contrary to industry standards relating to acceptable input power.
THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER REPRESENTATIONS AND WARRANTIES, EXPRESS OR IMPLIED; AND SELLER EXPRESSLY DISCLAIMS AND EXCLUDES ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. SELLER SHALL HAVE NO OBLIGATION OR LIABILITY TO REFUND ANY PORTION OF THE PURCHASE PRICE AND SHALL NOT BE LIABLE FOR ANY
SPECIAL, EXEMPLARY, INCIDENTAL, CONSEQUENTIAL OR PUNITIVE DAMAGES OR LOST PROFITS, OR DAMAGE TO PERSON OR INJURY IN CONNECTION WITH THE PURCHASE OR USE OF THE INSTRUMENT.
WORKING CLEARANCE ACCORDING TO TRANSMISSION FREQUENCY [M]
150 KHZ TO 80 MHZ
d= [1.2] √P
0.12
0.38
1.2
3.8
12
80 MHZ TO 800 MHZ
d= [1.2] √P
0.12
0.38
1.2
3.8
12
800 MHZ TO 2.5 GHZ
d= [2.3] √P
.023
0.73
2.3
7.3
23