ELECTROMAGNETIC ENVIRONMENT GUIDANCE

ELECTROMAGNETIC EMISSION

The Gemini EVO 810+980 soft tissue laser is intended for operation in the electromagnetic environment specified below. The customer or user of the Gemini EVO laser should make sure that it is used in such
an environment.
EMISSION TEST
RF emissions according to CISPR 11
RF emissions according to CISPR 11
Harmonic emissions according to IEC 61000-3-2
Voltage fluctuations/flicker emissions according to IEC 61000-3-3
INTERFERENCE IMMUNITY
The Gemini EVO laser is intended for operation in the electromagnetic environment specified below. The customer or user of the Gemini EVO laser should make sure that it is used in such an environment.
INTERFERENCE IMMUNITY TEST
IEC 60601-1-2 TEST LEVEL
Electrostatic discharge (ESD) according to
+/- 2, +/-4, +/- 6, +/-8 kV contact discharge
IEC 61000-4-2
+/- 2, +/-4, +/-8 kV, +/-15 kV air discharge
Electrical fast transient/burst according to
100kHz repetition +/- 2kV for power
IEC 61000-4-4
supply lines
100kHz repetition +/- 2kV for power input/
output lines
Surge voltages according to IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode voltage
Voltage dips, short interruptions and
Voltage (VAC) 100, 240 over 0% residual
variations of the power supply according to
with ½ cycles on a phase angle of
IEC 61000-4-11
0 ° / 90 ° / 180° / 270 °
Voltage (VAC) 100, 240 over 0% residual
with 1
cycles on a phase angle of
0 ° / 90 ° / 180° / 270 °
Voltage (VAC) 100, 240 over 70% residual
with 25
cycles on a phase angle of
0 ° / 90 ° / 180° / 270 °
Voltage (VAC) 100, 240 over 0% residual
with 250
cycles on a phase angle of
0 ° / 90 ° / 180° / 270 °
Magnetic field of power frequencies (50/60
3 A/m
Hz) according to IEC 61000-4-8
IMMUNITY TEST
IEC 60601 TEST LEVEL
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
18
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
The Gemini EVO laser uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to
GROUP 1
cause any interference in nearby electronic equipment.
CLASS B
The Gemini EVO laser is suitable for use in all establishments, including domestic establishments and those directly connected to
CLASS A
the public low-voltage power supply network that supplies buildings used for domestic purposes.
COMPLIES
COMPLIANCE LEVEL
+/- 2, +/-4, +/- 6, +/-8 kV contact discharge
+/- 2, +/-4, +/-8 kV, +/-15 kV air discharge
100kHz repetition +/- 2kV for power
supply lines
Not applicable
± 1 kV differential mode
± 2 kV common mode voltage
Voltage (VAC) 100, 240 over 0% residual
with ½ cycles on a phase angle of
0 ° / 90 ° / 180° / 270 °
Voltage (VAC) 100, 240 over 0% residual
with 1
cycles on a phase angle of
0 ° / 90 ° / 180° / 270 °
Voltage (VAC) 100, 240 over 70% residual
with 25
cycles on a phase angle of
0 ° / 90 ° / 180° / 270 °
Voltage (VAC) 100, 240 over 0% residual
with 250
cycles on a phase angle of
0 ° / 90 ° / 180° / 270 °
3 A/m
COMPLIANCE LEVEL
3 Vrms
3 V/m
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic
material, the relative humidity should be at least 50%.
The quality of the line power supply should be that of a typical commercial or hospital
environment.
The quality of the line power supply should be that of a typical commercial or hospital
environment.
The quality of the line power supply should be that of a typical commercial or hospital
environment.
If the user of the Gemini EVO laser requires it to continue functioning following
interruptions of the power supply, it is recommended to have the Gemini EVO laser
powered by an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels characteristic of a typical location in a
typical commercial or hospital environment
ELECTROMAGNETIC ENVIRONMENT GUIDANCE
Portable and mobile radio equipment must not be used within the recommended working
clearance from the Gemini EVO laser and its cables, which is calculated based on the
equation suitable for the relevant transmission frequency.
Recommended separation distance
d= [1.2] √P
d= [1.2] √P at 80 MHz to 800 MHz
d= [2.3] √P at 800 MHz to 2.5 GHz
where P is the nominal transmitter output in watts (W) specified by the transmitter
manufacturer and d is the recommended working clearance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey2 should be less than the compliance level3 in each frequency range.
Interference is possible in the vicinity of equipment bearing the following graphic symbol.