8930EN_6.fm 12/17/10 2:44 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)

Electromagnetic Compatibility Declaration (EN60601-1-2)

Table 6-6. Electromagnetic Emissions
The Model 8930 Radio Frequency Generator is intended for use in the electromagnetic
environment specified below. The customer or the user of the generator should ensure that it is
used in such an environment.
Emissions test
Radio-frequency
(RF) emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
EN 61000-3-2
Voltage
fluctuations/flicker
emissions
EN 61000-3-3
MA07445A008
Rev X
Compliance
Electromagnetic environment – guidance
The Model 8930 must emit electromagnetic energy
Group 2
in order to perform its intended function. Nearby
electronic equipment may be affected.
The Model 8930 is suitable for use in all
Class A
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
The device is suitable for use in all establishments,
Not Applicable
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
The Model 8930 is suitable for use in all
establishments, including domestic establishments
Class A
and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
The device is suitable for use in all establishments
Not Applicable
other than domestic and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Complies
The Model 8930 is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public
low-voltage power supply network that supplies
buildings used for domestic purposes.
Not Applicable
The device is suitable for use in all establishments
other than domestic and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Specifications
Device Specifications
2010-12 8930 System User Guide English
Medtronic Confidential
CRMRef_R06