Radio Frequency Generator Description
Table 2-7. Alert Messages (continued)
No.
Message
20
Temperature problem
Check hand piece
Contact Medtronic if
problem persists
21
Foot switch problem
Check foot switch or
remove foot switch and
use hand piece switch
25
Lesion stopped by user
User Interface
Explanation
Temperature measurement
from thermocouples is
°
outside 0
C to 150
range
Message clears when
temperature is within
range; displays for
minimum of ten seconds
Alarm is 2 short beeps
+ 1 long beep if radio
frequency on; if off,
2 short beeps
Optional foot switch is
stuck, or kept pressed for
more than five seconds
Foot switch should be
checked, or replace with a
new unit
Message stays on until foot
switch is released or
disconnected
Alarm is 2 short beeps
+ 1 long beep if radio
frequency on; if off,
2 short beeps
The clinician has stopped
radio frequency output
before end of lesion
If system conditions OK,
radio frequency generator
will switch to "System
ready" in approximately
five seconds
Message stays on until
system switches to
"System ready"
Alarm is 2 short beeps
+ 1 long beep
2010-12 8930 System User Guide English
Medtronic Confidential
CRMRef_R06
45
°