Device Specifications

Device Specifications

Radio Frequency Generator

English 8930 System User Guide 2010-12
MA07445A008
Rev X
Table 6-1. Model 8930 Radio Frequency Generator
a
Specifications
Description
Classification
Operating modes
RF output power
Impedance range
Impedance measurement
accuracy
System control
Temperature-sensor
range
Temperature accuracy
Urethral temperature
limits
Specification
IEC60601-1, Class I, Type BF, IPX0, no
sterilization, continuous use
This device complies with Part 18 of the
FCC rules.
This ISM device complies with Canadian
ICES-001.
Automatic dual needle
Automatic single needle
2 channels x 15 W (+/-10%) into 35–600
Ohms for Prostiva RF Therapy software
calibration–generator is rated at
2 channels x 50 W (+/-10%) into 20–600
Ohms
35-600 Ohms, nominal
+/- 10% + 5 Ohms for 50–200 Ohms
+/- 20% + 5 Ohms for 0–50 & 200–
999 Ohms
System only allows radio frequency energy
power output when the measured load
impedance is within 35 to 600 Ohms for one
or both channels
°
°
°
15
C to 140
C (59
F to 284
°
°
°
+/-2
C for 15
C to 60
C
°
°
°
(+/-4
F for 59
F to 140
F)
°
°
°
+/-3
C for 60
C to 140
C
°
°
°
(+/-5
F for 140
F to 284
F)
°
°
>43
C (109
F) caution tone sounds and
message displays
°
°
>47
C (117
F) radio frequency power
turns off, alert tone sounds, alert message
displays
Medtronic Confidential
CRMRef_R06
°
F)