Table 2-8. Caution Messages (continued)
No.
Message
58
Left needle off-impedance
(Ω) too high
May continue with right
needle
Needle may be in dry or
hard tissue, or outside
prostate
59
Right needle
off-impedance (Ω) too
high
May continue with left
needle
Needle may be in dry or
hard tissue, or outside
prostate
60
Left needle
off-temperature too low
May continue with right
needle
Needle may be near
bladder or blood vessel,
or outside prostate
Medtronic Confidential
CRMRef_R06
Explanation
Left needle is off due to
■
high impedance
(>600 ohms)
Left needle position is not
■
optimal, but treatment may
be continued with right
needle
Message stays on for ten
■
seconds
Alarm is 3 short beeps
■
with varied tones
Right needle is off due to
■
high impedance
(>600 ohms); treatment
may be continued with left
needle
Right needle position is not
■
optimal, but treatment may
be continued with left
needle
Message stays on for ten
■
seconds
Alarm is 3 short beeps
■
with varied tones
Left needle is off due to
■
low temperature; treatment
may be continued with
right needle
Left needle position is not
■
optimal, but treatment may
be continued with right
needle
Message stays on for ten
■
seconds
Alarm is 3 short beeps
■
with varied tones