Chapter 4
66

Patient and Hand Piece Preparation

Patient and Hand Piece Preparation

Attaching Return Electrode (Neutral Electrode)

text
English 8930 System User Guide 2010-12
MA07445A008
Rev X
Prepare the patient for the Prostiva RF Therapy procedure in
accordance with a standard cystoscopic procedure and the
following instructions. The instructions in the following sections
outline the attachment of the return electrode, draping of the
patient, and setting up of the hand piece. The configuration of the
radio frequency generator for patient treatment is outlined in
"Patient Treatment" on page 72.
Refer to the Model 8929 Hand Piece Instructions for Use for
detailed instructions, warnings, and precautions regarding the use
of the hand piece, return electrode, and tubing system.
Warning: Failure to observe the warnings outlined in the
Model 8929 Instructions for Use or failure to use the Model
8934 Split-pad Return Electrode can result in low system
output, failure of the system to function properly, or patient
injury (eg, burns).
1. Open the package and remove the Model 8934 Split-pad
Return Electrode.
2. Slowly remove plastic backing to expose
conductive/adhesive area without creasing the return
electrode surface (Figure 4-5). Discard the plastic backing.
Note: If the electrode is creased during removal, discard and
use a new return electrode.
Figure 4-5. Carefully Remove Return Electrode Backing.
Medtronic Confidential
CRMRef_R06