Herunterladen Inhalt Inhalt Diese Seite drucken

Medtronic ProstivaRF Therapy 8930 Bedienungsanleitung Seite 42

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 255
8930EN_2.fm 12/17/10 2:43 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)
Chapter 2
42
User Interface
English 8930 System User Guide 2010-12
MA07445A008
Rev X
Table 2-7. Alert Messages
No.
Message
11
Generator problem
Restart generator
Contact Medtronic if
problem persists
12
Generator overheated
Check air vents and wait 5
minutes before
continuing
Contact Medtronic if
problem persists
13
Hand piece not connected
Connect hand piece to
generator
Medtronic Confidential
CRMRef_R06
Explanation
Radio frequency generator
has a hardware or software
problem that may be
recoverable
Problem may be
correctable if user restarts
system
Message shows briefly
until system displays a
solid blue screen; this may
clear when system is
restarted
Alarm is 2 short beeps
+ 1 long beep if radio
frequency on; if off,
2 short beeps
Radio frequency generator
is overheated
Check bottom vents and
fan clearances; allow
generator to cool for at
least five minutes
Message stays on for five
minutes until generator is
cool enough to proceed
Alarm is 2 short beeps
+ 1 long beep if radio
frequency on; if off,
2 short beeps
Hand piece cable is not
properly connected to the
radio frequency generator
Hand piece must be
connected in order to
proceed
Message stays on until
user connects hand piece
cable
Alarm is 2 short beeps
+ 1 long beep if radio
frequency on; if off,
2 short beeps

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Medtronic ProstivaRF Therapy 8930

Inhaltsverzeichnis