Table 6-1. Model 8930 Radio Frequency Generator
a
Specifications
(continued)
Description
RF output power
frequency
Input power
Maximum RF voltage (per
channel)
Maximum RF current (per
channel)
Fuse protection
Case dimensions
Weight
Flammable anesthetic
mixture
Operating temperature
Temperature limitation
Humidity limitation
Power cable length
a
All measurements are approximate
Specifications
Device Specifications
Specification
473 kHz +/- 2 kHz sinusoidal, unmodulated
100-240 VAC, 200 VA, 50/60 Hz
173 Vrms (at 50 W, 600 Ohms-hardware
limit)
95 Vrms (at 15 W, 600 Ohms-software limit)
1.56 A (hardware limit)
0.66 A (software limit)
Two, T3.15A (5 mm x 200 mm), 250 V,
slow-blow fuses
36 cm W x 56 cm L x 15 cm H (closed) &
38 cm H (open) 14 in W x 22 in L x 6 in H
(closed)/15 in H (open)
9 kg (21 lb)
Not suitable for use in the presence of a
flammable anesthetic mixture with air,
oxygen, or nitrous oxide
°
°
°
10
C to 40
C (50
F to 104
°
°
°
-34
C to 57
C (-30
F to 135
°
°
5% to 95% RH 35
C (95
F)
3.0 m (10 ft)