Herunterladen Inhalt Inhalt Diese Seite drucken

Determining Number Of Treatment Planes - Medtronic ProstivaRF Therapy 8930 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 255
8930EN_4.fm 12/17/10 2:44 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)
Chapter 4
64
Patient Assessment

Determining Number of Treatment Planes

English 8930 System User Guide 2010-12
MA07445A008
Rev X
The following table (Table 4-2) provides guidelines for determining
the number of treatment planes based on the distance from the
bladder neck to the verumontanum. Ideally, a minimum of two
planes should be treated providing that the distance from the
needle placement to the bladder neck and from the needle
placement to the verumontanum remains 0.75-1.0 cm.
A treatment plane consists of delivery of energy to the right and
left lobes at the same level as shown in Figure 4-3.
Note: For each additional centimeter in length, add an additional
treatment plane.
The determination of the number of treatment planes is in the
clinician's sole medical judgment.
Table 4-2. Determining Number of Treatment Planes from Step 1
Measurement
Bladder Neck to
Number of Treatment
Verumontanum
Length
<3 cm
3-4 cm
4-5 cm
Figure 4-3. Example of 4 Completed Treatment Planes (8 Lesions)
Medtronic Confidential
Treatment Planes
Planes
2-3
proximal and distal
3-4
proximal
mid-point and distal
4-5
proximal
mid-point
mid-point and distal
CRMRef_R06