Herunterladen Inhalt Inhalt Diese Seite drucken

Medtronic ProstivaRF Therapy 8930 Bedienungsanleitung Seite 46

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 255
8930EN_2.fm 12/17/10 2:43 pm
UCxxxxxx EN
7 x 9 inches (178 mm x 229 mm)
Chapter 2
46
User Interface
English 8930 System User Guide 2010-12
MA07445A008
Rev X
Table 2-7. Alert Messages (continued)
No.
Message
31
Needle impedance (Ω) too
high
Retract needles, and
reposition or reduce
length
Needle may be in dry or
hard tissue, or outside
prostate
32
Needle temperature too
high
33
Needle temperature too
low
Retract needles, and
reposition or reduce
length
Needle may be near
bladder or blood vessel,
or outside prostate
34
Urethra temperature too
high
Medtronic Confidential
CRMRef_R06
Explanation
Impedance exceeds
600 ohms; stops lesion
Needle position is not
optimal, and it must be
changed
Message stays on for ten
seconds
Alarm is 2 short beeps
+ 1 long beep
Radio frequency delivery
has raised needle
°
temperature above 125
C;
stops lesion
Message stays on for ten
seconds
Alarm is 2 short beeps
+ 1 long beep
Needle position is not
optimal, and it must be
changed
Needle may be incorrectly
positioned in bladder;
stops lesion
Needle position or length is
not optimal, and it must be
changed
Message stays on for ten
seconds
Alarm is 2 short beeps
+ 1 long beep
Radio frequency delivery
has raised urethral
°
temperature above 47
C;
stops lesion
Clinician must irrigate
more frequently to help
prevent the urethra from
overheating
Message stays on for ten
seconds
Alarm is 2 short beeps
+ 1 long beep

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Medtronic ProstivaRF Therapy 8930

Inhaltsverzeichnis