Table 2-6. Status Messages (continued)
No.
Message
4
Lesion in progress with
left needle only...
5
Lesion in progress with
right needle only...
6
Lesion complete
Retract and reposition
needles
Start new lesion
Medtronic Confidential
CRMRef_R06
Explanation
Lesion treatment is with left
needle only because right
needle may be outside
prostate and has shut off
Message stays during
lesion unless caution
message displays
Alarm is 1 short beep
every 4 seconds
Both blue LEDs are on
solid
Lesion treatment is with
right needle only because
left needle may be outside
prostate and has shut off
Message stays on during
lesion unless caution
message displays
Alarm is 1 short beep
every 4 seconds
Both blue LEDs are on
solid
Lesion time of 2.3 minutes
has expired and system is
ready to start next lesion
Message stays on until
clinician starts next lesion
Alarm is 1 long beep
Both blue LEDs flash at a
continuous rate